1. Setup
ABL90 FLEX reference manual
1-42
Locking/unlocking a parameter:
Step Action
1.
Ensure that the analyzer is not connected to the RADIANCE system,
as parameters can be locked/unlocked from the RADIANCE system.
2.
Highlight a parameter on the screen, using the scroll facilities.
3.
Press Lock/Unlock. (This button is grayed-out if the analyzer is
connected to the RADIANCE system.)
4.
To unlock a parameter, highlight it and press Lock/Unlock. The
traffic light on Analyzer status will change from YELLOW to a color
corresponding to the analyzer's overall status.
NOTICE: A locked parameter will show YELLOW on the parameter bar and will
change the overall analyzer status traffic light on the Analyzer status screen to
YELLOW. The parameter value will be absent from the printout; however, the
locked parameter will be calibrated.
Editing the parameter setup:
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...