6. User-defined corrections
ABL90 FLEX reference manual
6-8
Correction factors for electrolyte and metabolite
parameters
The following corrections to the slope and the offset are possible within the
stated limits for:
Blood samples:
Parameter Slope
Offset
cK
+
0.80-1.20
1.0 mmol/L
cNa
+
0.80-1.20
10 mmol/L
cCa
2+
0.80-1.20
1.00 mmol/L
cCl
–
0.80-1.20
10 mmol/L
cGlu 0.750-1.250
5.0 mmol/L
cLac 0.750-1.250
5.0 mmol/L
The Radiometer default values for the electrolyte and metabolite parameters
must be reset manually by the user on the Parameter setup screen to 1.000
for each parameter.
Correcting
slope and
offset
Resetting
corrections to
default values
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...