5. Sensors and optical system
ABL90 FLEX reference manual
5-44
Calibration of the optical system
The optical system is calibrated at two points using the following:
The S7770 ctHb Calibration Solution with a known dye concentration to
determine the cuvette path length,
l.
A transparent solution from the solution pack in the analyzer to determine the
zero point,
I
o
.
The zero point,
I
o
, is the current (or intensity) measured by the photodiode
array on the transparent solution in the cuvette. During this "blank calibration"
the
ctHb is calibrated to this zero point.
I
o
is measured automatically during system start up and during calibrations.
The cuvette path length (i.e. the length of the light path) is determined from
Lambert-Beer’s Law by measuring the absorbance of the colored dye present in
the tHb Calibration Solution (S7770), which has a known equivalent hemoglobin
concentration.
Beer’s Law:
A =
ε
x
C
dye
x
l
where:
A
= absorbance
ε
= extinction coefficient
C
dye
= concentration of colored dye
l
= length of light path
It is recommended that a tHb calibration is performed every three months. See
section
tHb calibration in chapter 6: Calibration in the ABL90 FLEX operator's
manual for further information about the tHb calibration.
Calibration
materials
Zero point
Cuvette path
length
tHb calibration
frequency
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...