10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-36
No.
Message
Interpretation
Operator action
1195 ABL not
connected to QA
Portal - incorrect
password
The analyzer was refused
connection to the QA Portal
due to incorrect password.
- Enter the correct
password in the
analyzer's QA Portal
Communication Setup.
1196 ABL connected to
QA Portal
The analyzer is connected to
the QA Portal.
- No action required. For
information only.
1197 ABL disconnected
form QA Portal
The analyzer is disconnected
from the QA Portal.
- No action required. For
information only.
1198 ABL<>QA Portal
communication
error - XML
packet could not
be parsed
Communication error
between the analyzer and
the QA Portal.
- Contact IT engineer or
Radiometer service
representative.
1199 FTC program has
been retried
This message is found in the
Activity Log when a
measurement or calibration
activity has been retried due
to error.
- No action required. For
information only.
1200 Solution pack
empty
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1201 Solution pack
lifetime expired
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1202 Expiration date
reached
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1203 Lifetime in
analyzer
exceeded
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1204 No more
activities left
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...