ABL90 FLEX reference manual
5. Sensors and optical system
5-31
Measurement - pO
2
On whole blood, pO
2
is adjusted with the sensitivity value and the measured pO
2
is therefore determined as follows:
2
2
O (meas)
O (sens,adjusted)
S
p
p
The measured value is applied a 2
nd
order blood correction, to compensate for
the varying buffer value of blood, as a function of pO
2
tension. A second-order
correction is applied:
2
2
1
2
2
2
3
O (display) k O
k O + k
p
p
p
NOTICE
:
During a measurement, the sensor technology used offers detection of
any air bubble in front of the
pO
2
sensor that could lead to significant errors.
The following parameters are range checked:
Sensitivity
Sensitivity drift
Status
Checks
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...