10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-26
No.
Message
Interpretation
Operator action
1031 Oxi initialization
in progress
Oxi initialization in progress. - Please wait up to 50
minutes before
restarting the analyzer.
- Restart the analyzer.
- Contact Radiometer
service representative.
1032 Oxi data
collection
problem
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1033 Oxi task was not
finished
Internal software problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1034 Oxi hardware
problem
An Oxi hardware problem
has occurred.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1045 Unable to read
consumable
information
Unable to read information
stored on either sensor
cassette or solution pack.
- Reinstall the solution
pack and sensor
cassette.
- Restart the analyzer.
- Contact Radiometer
service representative.
1061 Pressure test
flow error
The sample transport
through the analyzer is
hindered.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1062 Pressure test
pressure error
A leak has been found in the
solution transport.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1063 Pressure test
vacuum error
A leak has been found in the
solution transport.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1064 Temperature in
sensor cassette
top out of range
Hardware temperature error. - Ensure that the
ambient temperature is
between 15 and 32 °C.
- If the system has just
performed a cold start,
wait for the error to
disappear.
- Replace the fan filter,
if dirty.
- Shield the analyzer
from direct sunlight and
other heat sources.
- Contact Radiometer
service representative.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...