ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-17
No.
Message
Interpretation
Operator action
773
Remote operator
logged on with
user:
A remote operator has
logged on the analyzer via
NetOp.
- No action required.
774
Remote operator
logged off with
user:
An operator, remotely logged
on via NetOp, has logged off,
or has been logged off by a
local operator.
- No action required.
775
Failed to restore
Default Setup
Restoring analyzer setup to
default values has failed.
- Contact Radiometer
service representative.
776
Succeeded to
restore Default
Setup
Restoring setup to default
values is completed.
- No action required.
780
RADIANCE
communication
enabled
RADIANCE communication
has been enabled as part of
the RADIANCE Connection
Setup.
- No action required. For
information only.
781
RADIANCE
communication
disabled
RADIANCE communication
has been disabled as part of
the RADIANCE Connection
Setup.
- No action required. For
information only.
782
RADIANCE
output queue
cleared
The output queue was
cleared in the RADIANCE
Connection Setup.
- No action required. For
information only.
783
Automatic
backup started
Automatic backup (selected
in Disk Functions Setup) has
started.
- No action required. For
information only.
785
Automatic
archiving started
Automatic archiving (selected
in Disk Functions Setup) has
started.
- No action required. For
information only.
786
Automatic
archiving
completed
Automatic archiving (selected
in Disk Functions Setup)
completed successfully.
- No action required. For
information only.
787
Export of data
logs started
Export of data logs was
started by the user.
- No action required. For
information only.
798
User logged on
User logged on successfully. - No action required. For
information only.
799
User logged off
User logged off.
- No action required. For
information only.
800
Logon attempt
failed
User tried to log on but did
not provide a valid password.
- Provide a valid
password to log on.
810
pH locked
The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...