10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-32
No.
Message
Interpretation
Operator action
1147 Inlet opened
during rinse
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1148 Inlet open during
calibration
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1149 Inlet open during
wet section
activity
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1150 Inlet closed
without
aspirating
sample
This message is shown in the
Activity Log when a
measurement has been
cancelled due to inlet being
closed before aspiration
could be completed.
- No action required. For
information only.
1151 Inlet not closed:
no sample
aspirated
This message is shown in the
Activity Log when a
measurement has been
cancelled due to inlet being
closed too late.
- No action required. For
information only.
1152 The solution pack
chip data could
not be read or
written during
replacement
This message is shown in the
Activity Log when a
replacement of the sensor
cassette or solution pack has
failed. The reason was that it
was impossible to
communicate with the chip
on the consumable.
- Repeat replacement
operation.
1157 No valid FTC
programs
detected
System error.
- Contact Radiometer
service representative.
1165 Solution pack not
properly installed
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
1166 Solution pack
expired
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...