10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-20
No.
Message
Interpretation
Operator action
826
HbF locked
The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
827
tBil locked
The parameter has been
locked by a RADIANCE
operator, as reflected in the
Activity Log.
When a parameter is locked,
presumably due to problems
with QC, the parameter is
repressed in patient results.
- Await corrective
actions initiated by the
RADIANCE operator.
Removal condition:
- Determined by the
RADIANCE operator.
831
pH unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
832
pCO
2
unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
833
pO
2
unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
834
K unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
835
Na unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
836
Cl unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
837
Ca unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
839
Glu unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
840
Lac unlocked
The message is used in the
Activity Log to indicate that a
previously locked parameter
has been unlocked.
- No action required. For
information only.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...