ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-11
No.
Message
Interpretation
Operator action
529
Inlet LS failed to
calibrate
Inlet liquid sensor failed to
calibrate.
- Repeat the liquid
sensor calibration.
- Contact Radiometer
service representative.
531
Sensors LS failed
to calibrate
Liquid sensor near the sensor
cassette failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace if
necessary.
- Contact Radiometer
service representative.
537
OXI LS failed to
calibrate
OXI module liquid sensor
failed to calibrate.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
581
OXI spectrum
mismatch
Spectrum deviates from the
expected blood or QC
spectrum. Measurement may
be unreliable.
- Check the patient
record for medication
containing possible
interfering substances.
- Start a calibration.
- Contact Radiometer
service representative.
582
tHb calibration
cuvette factor
outside limits
tHb calibration failed.
- Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- Successful tHb
calibration.
584
tHb calibration
wavelength
outside limits
tHb calibration failed.
- Perform a calibration.
- Repeat the tHb
calibration.
- Contact Radiometer
service representative.
Removal condition:
- A successful tHb
calibration
588
Measured QC
value lower than
statistical range
The parameter value is below
the lower limit of the user-
defined statistical range.
Measurement is not included
in statistics.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference manual
for details on the
evaluation of the
results.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...