ABL90 FLEX reference manual
6. User-defined corrections
6-7
The following recommendations apply to ctBil:
Item Description
Units
mol/L
Sample
Radiometer recommends that human plasma or serum is
used with pH
7.4 (the analyzer reading). Zero point
sample could be adult sample (ctBil
0
mol/L) and
maximum point could be an unconjugated bilirubin sample
(neonatal) with ctBil
300-400
mol/L.
Averaging repeated measurements on samples from
different donors gives an optimized accuracy of the
correction. Averaging repeated measurements on samples
from the same donor also improves the accuracy.
Commercial bilirubin standards can interfere with bilirubin
measurement because they may have an absorbance
spectrum different from that of human plasma.
Slope 0.500-1.500
Offset
100
ctBil
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...