ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-25
No.
Message
Interpretation
Operator action
1018 Oxi electronic
adjustment error
Oxi hardware problem.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1019 Oxi Blank Cal.
outside limits
Peak value of Blank Cal.
spectrum intensity is outside
acceptance limits.
- Check solution pack.
The cuvette must be
filled with liquid during
Blank calibration.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1020 Oxi neon
intensity outside
limits
Oxi hardware problem.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1021 Oxi neon
correction
outside limits
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1022 Oxi background
correction
outside limits
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1023 Oxi spectrometer
memory read
problem
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1024 Oxi spectrometer
memory write
problem
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1025 Oxi hemolyzer
tuning problem
Oxi hardware problem.
- Restart the analyzer
- Contact Radiometer
service representative.
1026 Oxi hemolyzer
frequency
problem
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1027 Oxi hemolyzer
temperature
deviation too
high
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1028 Oxi neon voltage
outside limits
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1029 Oxi light source
voltage outside
limits
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
1030 Oxi hemolyzer
voltage outside
limits
Oxi hardware problem.
- Restart the analyzer.
- Contact Radiometer
service representative.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...