Note to the users of the ABL90 FLEX analyzer
This note to users outlines the most significant changes and improvements of software
version 2.4 in the English version of the ABL90 FLEX reference manual (994-527).
Changes/Description
CHAPTER 3:
Calibration
pCO
2
, cGlu, cLac calibration
With the warm start feature in software version 2.4, the last
sentence in this section: When a new sensor cassette is installed,
the sensitivity calibration is performed more frequently" is not
necessarily true and should, therefore, be ignored. The text
should, therefore, be as follows:
pCO
2
, cGlu, cLac are sensitivity calibrated on CAL3 and status
calibrated on CAL1. For further information on the pCO
2
, cGlu,
cLac calibration, see Calibration of the pCO
2
sensor in section
pCO
2
sensor and Calibration of the metabolite sensors, in section
Metabolite sensors in chapter 5 in this manual. pCO
2
, cGlu, cLac
are sensitivity calibrated every four hour and status calibrated
with every measurement.
CHAPTER 5:
Calibration
Sensitivity, Definition.
In the table with the sensitivity range limits, the pO
2
max %
value has been changed from 105 to 110%.
The sensitivities are range checked:
pH
pCO
2
pO
2
cK
+
cNa
+
cCa
2+
cCl
–
cGlu
cLac
% % % % % % % pA/mmol/L
pA/mmol/L
Min.
85 60 85 85 85 85 75 100
100
Max. 105 105 110 105 105 105 105 2000
2000
CHAPTER 7:
Interference tests
For ClO
4
–
,
interference on cK
+
(4 mmol/L level), cCa
2+
(1.25
mmol/L level), and cCl
(110 mmol/L level) has been detected for
test concentration 1.5 mM and the other test concentration levels
should be ignored. The new interference results are, thereafter,
as follows:
Interference on....
Substance Test
con-
centration
cK
+
(4 mM
level)
cNa
+
(140
mM
level)
cCa
2+
(1.25
mM
level)
cCl
-
(105
mM
level)
Test
matrix
Perchlorate
(ClO
4-
)
1.5 mM
0.3
-
0.27*
4-30
Plasma
* Depending on the pH level
A "-" indicates that interference has not been measured on the respective parameter.
Introduction
Brief overview
of the change
© 2010 Radiometer Medical ApS. All Rights Reserved.
995-675. 201006A.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...