10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-4
No.
Message
Interpretation
Operator action
94
Value below
reportable range
The parameter value is below
the reportable range.
- Check for and remedy
other errors related to
the result, system
messages or calibration
status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform a
measurement on new
blood sample.
117
LIS/HIS: Invalid
connection
configuration
The communication
configuration or the protocol
definition was invalid.
- Check the
communication
parameters specified in
Communications Setup.
128
LIS/HIS: Failed
to open
connection
The communication hardware
was busy or the remote
system did not respond.
- Check that the remote
system is running,
correctly configured and
responding.
- Check communication
parameters, e.g. baud
rate, parity, IP address,
etc., as defined in
Communication Setup.
- Reboot the analyzer.
129
LIS/HIS: Failed
to close
connection
Messages were queued when
the communication channel
was closed. Results and
other messages sent by the
analyzer to a remote system
may be lost.
- If the problem
persists, check the
communication
hardware. The remote
system may lack buffer
capacity.
131
LIS/HIS: Failed
to send packet
A communication error
occurred while sending a
message. The message was
not sent.
- Check that the remote
system is running and
responding.
- Check the
communication
hardware, including
cables.
- Repeat sending.
132
LIS/HIS: Failed
to receive packet
An error occurred while
receiving a message. The
analyzer was not able to
recognize the received
massage.
- Check that protocol
types are correctly
configured on both the
analyzer and the remote
system.
- Contact Radiometer
service representative.
133
LIS/HIS:
Connection lost
A previously established
LIS/HIS connection has been
lost.
- Check that the remote
system is running and
responding.
- Check cables.
134
LIS/HIS:
Connection
established
The connection was
successfully established.
- No action required. For
information only.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...