5. Sensors and optical system
5-49
The uncorrected hemoglobin concentration,
ctHb(sample), measured on
capillary or syringe samples is corrected as follows:
cuv
tHb(sample)
tHb(sample,corr)
F
c
c
where:
ctHb(sample,corr)
= corrected
ctHb
F
cuv
= analyzer-dependent cuvette path length constant
determined at tHb calibrations and automatically
stored by the analyzer
The uncorrected total bilirubin concentration,
ctBil(sample), measured on
capillary or syringe samples is corrected as follows:
cuv
tBil(sample)
tBil(sample,corr)
F
c
c
F
cuv
is the same as for tHb.
Corrections for
ctHb
Corrections for
ctBil
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...