ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-3
No.
Message
Interpretation
Operator action
1
Inconsistent
software
versions. Please
contact service
Inconsistent software
versions for different
modules. May appear after
replacing a complete module
or as a result of an
incomplete software
upgrade.
- Contact Radiometer
service representative.
Removal condition:
- Successful software
consistency check.
83
Value above
reference range
The parameter value is
above the user-defined
reference range.
This is only a message, not
an error.
- No action required.
84
Value below
reference range
The parameter value is below
the user-defined reference
range.
This is only a message, not
an error.
- No action required.
85
Value below
critical limit
The parameter value is below
the user-defined critical limit.
This is only a message, not
an error
- No action required.
86
Value above
critical limit
The parameter value is
above the user-defined
critical limit.
This is only a message, not
an error.
- No action required.
89
Measured QC
value above
control range
The measured parameter
value is above the control
range.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference
manual.*
90
Measured QC
value below
control range
The measured parameter
value is below the control
range.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference
manual.*
93
Value above
reportable range
The parameter value is
above the reportable range.
- Check for and remedy
other errors related to
the result, system
messages or calibration
status.
- Perform QC. If the QC
result is accepted, the
blood sample may be
suspected.
- Perform measurement
on new blood sample.
* The ABL90 FLEX reference manual includes a quality control appendix for
manual QC.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...