7. Performance characteristics
ABL90 FLEX reference manual
7-4
Parameter
Primary Reference Method
Secondary
Reference
Method
Reference
c
Glu Spectrophotometry,
using
the
hexokinase (HK) method
recommended by NCCLS, measured
on serum.
N/A
[Ref. 5]
c
Lac
Spectrophotometry using a lactate
dehydrogenase (LDH) method,
measured on serum.
N/A
[Ref. 8]
c
tHb
HiCN method recommended by
NCCLS.
ABL735 [Ref.
6]
s
O
2
Tonometry:
100%: whole blood is tonometered
with a gas mixture containing
94.4% O
2
and 5.6% CO
2
. 0%:
whole blood is tonometered with a
gas mixture containing 94.4% N
2
and 5.6% CO
2
+ dithionite.
ABL735
F
O
2
Hb
Measured according to the following
relation:
F
O
2
Hb = 1 - (FHHb + FCOHb +
F
MetHb)
ABL735
F
HHb
0%: whole blood is tonometered
with a gas mixture containing
94.4% N
2
and 5.6% CO
2
+
dithionite.
ABL735
F
COHb
Gas chromatography: The
Standards are carbon monoxide
mixtures with atmospheric air,
whose purity is validated in
accordance with NIST SRM 1678 (50
ppm CO in N
2
)
ABL735
F
MetHb
Spectrometry, modified Evelyn-
Malloy method.
ABL735 [Ref.
7]
c
tBil
The reference method for total
bilirubin is a spectrophotometric
method (wet chemistry based on a
method from Bayer Healthcare,
Tarrytown USA).
The method is traceable to NIST
SRM916a Bilirubin.
ABL735
F
HbF
The reference method is based on
Cation Exchange HPLC.
ABL735 [Ref.
15]
General reference: [Ref. 10].
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...