10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-22
No.
Message
Interpretation
Operator action
963
Leak currents
detected
Leak currents were detected
during system calibration and
may distort measuring
results.
- Replace liquid- or
sensor cassette.
- Contact Radiometer
service representative.
964
Leak currents
detected
Leak currents were detected
during system calibration and
may distort measuring
results.
- Check sensor cassette
status and replace, if
necessary.
- Contact Radiometer
service representative.
970
Replace solution
pack
This message is shown when
the solution pack needs to be
replaced. The analyzer will
enter "User-intervention
required".
- Replace solution pack.
971
Replace sensor
cassette
This message is shown when
the sensor cassette needs to
be replaced. The analyzer
will enter "User-intervention
required".
- Replace sensor
cassette.
972
Sensor
conditioning due
The sensor cassette must
soon be replaced. A new
sensor cassette should
preferably be conditioned
now.
- Start conditioning a
new sensor cassette.
973
Printer paper
must be replaced
No more paper in printer.
- Insert new printer
paper.
978
Flow selector
calibration error
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
979
Inhomogeneous
rinse solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
980
Inhomogeneous
QC1 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
981
Inhomogeneous
QC2 solution
Shown in the Activity Log
when "User-intervention
required" has been entered
due to this reason.
- The analyzer will
automatically enter
"User-intervention
required". Follow the
instructions shown on
the screen.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...