1. Setup
ABL90 FLEX reference manual
1-14
Step Action
1.
Highlight a parameter on the Reference ranges and critical
limits screen, using the up/down arrows.
2.
Enter sample type, sex and age group, if required.
3.
Press Edit to edit entries for a highlighted parameter.
4.
If any entries are present and you cannot use any of them, press
Clear limits.
Then enter new critical and reference limits, using the keypad and
confirming each entry with Enter.
5.
To change a value, touch and highlight it. Then enter the limit and
confirm with Enter.
6.
Press Back to return to the Reference ranges and critical limits
screen.
7.
Highlight a parameter in the "Parameter list" box to view the limits
on the Reference ranges and critical limits screen.
Setting
reference and
critical limits
for each
parameter
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...