ABL90 FLEX reference manual
1. Setup
1-41
Parameters and input setup
Press Menu > Utilities > Setup > General setup > Parameters and input
setup and activate a button to enter a program.
The following programs are available:
Parameters
Units
User-defined data items
User-defined notes
Parameter setup
In this program you can do the following:
Disable or enable a parameter
Repress parameters if problems are detected
Lock a parameter
Make user-defined corrections for each measured parameter
Make out-of-range suppression of oximetry parameters and bilirubin
Disabling/enabling a parameter:
Step Action
1.
Highlight a parameter on the screen, using the scroll facilities.
2.
Press Enable/Disable to include/exclude the parameter from a
parameter profile and the parameter bar. Note that pH, pCO
2
and
pO
2
cannot be excluded.
Program
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...