Levelling
At the end of the ventilation phase, the chamber is brought back to atmospheric pressure (
LEVELLING
) by injecting sterile outside air to allow the
opening of the door and the retrieval of the load:
1
1 3 4 ° C P O R O U S
L E V E L L I N G
8 6 . 9 ° C
- 0 . 4 3 b a r 2 9 : 2 1
CHAMBER
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
Levelling
(pressure)
from -0,45 to –0,10 bar
(-0,10 bar to the value of 0 bar in the
memory)
OFF OFF ON OFF OFF OFF OFF OFF OFF
Levelling
(waiting time for the
release of the locking
door mechanism)
Waiting time = 15”
(counting start from pressure–0,10 bar)
OFF OFF ON OFF OFF OFF OFF OFF OFF