When the wait phase ends, the pressure verification phase, true and proper, begins (
LEAKAGE PERIOD
), with a duration of 10 minutes:
V A C U U M T E S T
L E A K A G E P E R I O D
1 0 : 0 0
- 0 . 7 9 b a r 0 6 : 4 8
In this phase, a variation of
up to ±0.02 bar
is allowed, compared to the initial phase value.
Higher variations cause the test to fail. The time is counted down until the phase is completed.
When this phase is also completed, the pressure is brought back to atmospheric pressure.
V A C U U M T E S T
L E V E L L I N G
- 0 . 2 9 b a r 1 7 : 1 9
When the program finishes, the display will read:
V A C U U M T E S T
T E S T P A S S E D
- 0 . 0 1 b a r 1 7 : 4 4
The end of the program is signaled with a beep.
e
If the pressure change exceeds the pre-set limit, the program is interrupted and alarm message is generated. See a complete description
of the alarms in
Problem solution
.
The duration of the test is about
18 minutes
.
When the door is opened at the end of the program, a report of the test cycle is printed with all the salient data.
For complete details about printed reports, refer to
Print report examples
.
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...