MIN AND MAX WATER LEVEL PROBES (DISTILLED WATER
TANK)
432#00050000
MAX WATER LEVEL PROBE (USED WATER TANK)
432#00060000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1.
Remove the covers (see card
Gr7-1
);
2.
Remove the wiring of the water level probe from the CPU board;
3.
Maintaining steady the probe body, loosen the ring nut and remove the
probe;
4.
Mount the new probe, assembly all items and restore connections proceeding
in reverse order as above;
5.
Run a sterilization cycle.
FLOAT MAX
(RECOVERY RESERVOIR)
FLOAT MAX
(MAIN RESERVOIR
FLOAT MIN
(MAIN RESERVOIR