
Rev. 1
Page 3-
6
which is automatically transformed to the message:
( A l a r m M e s s a g e )
R E S E T S Y S T E M
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Perform a
RESET
(described below) and then turn-off the device and check the alarm.
e
Alarms that intervene outside of a program do not produce a printed report
.
RESETTING THE SYSTEM
The system is
RESET
in two
alternative
ways, depending on the alarm that occurred:
1. Press
the
PROGRAM SELECTION
key for about 3 seconds.
A beep confirms the RESET;
Never turn the device off before executing a RESET.
2.
Turn-off the device and then power-on using the main switch
.
Upon power-up, the sterilizer will perform its normal initial test
.
After RESET, and any technical intervention necessary to eliminate the fault, the device will go to STAND-BY mode, ready to execute a new
program.
>3s
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...