VACUUM DOUBLE PUMP ASSEMBLY
A#0BM3100000
VACUUM DOUBLE PUMP
400#00290000
CENTRAL GASKET
400#00150000
MEMBRANE
400#00240000
“VITON” SHUTTERS
400#00180000
VACUUM PUMP HEAD
400#00120000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1.
Remove the covers (see card
Gr7-1
);
2.
Remove, from the bottom frame, the screws fixing the vacuum pump;
3.
Remove the sheathed pipes and the wiring from the pump;
4.
Remove the pump and disassembly it as necessary (see figure);
5.
replace the part, restore connections and assembly all items proceeding in
reverse order as above;
6.
Perform a vacuum test;
7.
Run a sterilization cycle.