Rev. 1
Card Gr7-
3
FRONT FRAME ASSEMBLY
A#0BP3000000
PLASTIC FRONT FRAME
C#3BP1830000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove plastic door cover (see card
Gr7-2
);
3. Remove the printer unit (see card
Gr1-20
);
4. Cut the nylon clamp on the mains switch wiring and disconnect it from the
switch;
5. Remove the LCD and keyboard connections from the CPU board;
6. Remove the bacteriologic filter;
7. Remove left and right rails;
8. Unscrew the screws fixing the plastic front cover; the screws are found rear
the front frame and printer compartment;
9. Unlock the four tabs of the plastic cover from the front frame slots around the
chamber circumference;
10. Carefully move left side the front plastic cover and remove it though the door;
11. Remove labels, LCD, keyboard, mains switch, plastic cap from the front
cover and mount them on the new cover;
12. Mount the new front cover, assembly all items and restore connections
proceeding in reverse order as above;
13. Perform a sterilization cycle.
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...