Rev. 1
Card Gr4-
1
WATER TANK ASSEMBLY
A#0BP2900000
WATER TANK
C#3BP1700000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1.
Remove the covers (see card
Gr7-1
);
2.
Empty both reservoirs;
3.
Remove the wiring from the water level probes and the rear pipes from the
reservoirs;
4.
Remove the right (1) and left (2) rail;
5.
Remove screws (3) fixing the reservoir;
6.
Lift the reservoir (4) and remove the bottom pipes;
7.
Mount the new reservoir, assembly all items and restore connections
proceeding in reverse order as above;
8.
Run a sterilization cycle.