Rev. 1
Page 3-
57
H 150
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
MPX BROKEN
Pressure transducer
broken
Pressure reading by the
transducer higher than 2.35
bar
Fault in the wiring connecting the pressure
transducer to board
Fault in the pressure transducer
Steam leakage from the pressure
transducer
Replace the pressure transducer - see card Gr1-4
Pressure transducer uncalibrated
Send the sterilizer to the Mocom Service department