ATTACHMENT G -
START-UP STEAM GENERATOR
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
Perform this procedure when:
a. The sterilizer remains unused for more one month;
b. The sterilizer was stored at a temperature below 0°C;
c. Water pump is replaced;
d. Steam generator is replaced;
e. Alarm concerning water absence in the reservoir is occurring.
Procedure
1. Remove the covers (see card
Gr7-1
);
2. Switch on the unit and enter the SETUP mode, SERVICE menu, type the
access code “++--++--“, and select the option H2O CIRCUIT;
3. Press the Start/Stop key and check the operation of the water pump and that
the water flows into the chamber;
4. Wipe the chamber and exit the SETUP mode;
5. Perform a sterilization cycle.
N.B. In case of problem, refer to the repairing card of the malfunctioning
component.