LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
OVERHEATING
PT2
PT2 overheating
PT2 detects a temperature
value greater than 230°C
Fault in the board
Replace the CPU board- see card Gr1-1
Fault in the pressure transducer
Replace the pressure transducer - see card Gr1-4
Steam generator clogged
Replace the lower section of the steam generator
- see card Gr4-6
Stored calibration values changed
Send the sterilizer to the Mocom Service department