Rev. 1
Attach.
18
ATTACHMENT O -
TEST THE UNIT THROUGH CONTINUOUS CYCLES
Use this procedure in case you need to test continuously the sterilizer for a lot of
time and be sure that this test will be normally concluded.
WARNING
The sterilizer is not normally provided with a water filling device and the number
of cycles is so limited.
1. Connect an external tank for the automatic water filling;
2. Connect an external tank for recovering the steam discharge at the end of the
cycle;
3. Fill the external tank with distilled water;
4. Enter SETUP mode, ADVANCED menu and set the FILLING OPTIONS on
the item AUTOMATIC;
5. Move to SERVICE menu and TEST CYCLES, select the option desired;
6. Exit the SETUP mode;
7. Select a cycle and enter the Start command.
WARNING
At the end of the test, reenter the SETUP menu, SERVICE menu, TEST CYCLES
and select the option “TEST CYCLES OFF” in order to set the sterilizer in
standard mode.
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...