LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
One or both the steam generator cartridges
failed
Check the steam generator cartridge.
Replace the failed cartridge – see card Gr1-19.
Band heater in short-circuit.
Check the band heater.
replace the band heater – see card Gr1-18.
Damage during the equipment transport.
Replace the damaged part.