Rev.1
Page 1-
41
ACTIVATING CONFIGURATION OPTIONS – SPECIAL MENU
Acquisition of the ambient pressure
(AMBIENT PRESSURE on the SPECIAL menu)
The first time the sterilizer is used and after any reinstallation, the sterilizer must acquire the ambient pressure.
This operation is
necessary
or the correct operation of several of the device's auxiliary systems.
When
AMBIENT PRESSURE
is activated, the following screen appears:
A C Q U I S I T I O N O F T H E
A M B I E N T P R E S S U R E
↵
t o e n t e r
t o e x i t
e
Check that the sterilizer door is completely OPEN. If you try to acquire the pressure with the door closed the following message will be
displayed:
O P E N T H E D O O R
T O C O N T I N U E
which remains until the door is opened. Confirm the acquisition of the data by pressing the
↵
key. This message appears:
P R E S S U R E V A L U E
S E T I N T O M E M O R Y
accompanied by a beep. The ambient data pressure has been acquired.
On the other hand, press the
↑
key to cancel the operation.
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...