Rev. 1
Attach.
9
ATTACHMENT H -
RELEASE DOOR LOCKING MECHANISM VIA SOFTWARE-
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove the jumper X21 of the CPU board;
3. Press the hidden key on the control panel, switch-on the unit and release the
key only when the message “
LOCKING DEVICE
” appears on the LCD
display;
4. Press the Start/Stop key to release the door locking mechanism;
5. Open the door, press the arrow-up key
to exit this status, and start the
Self test;
6.
GAM type pc boards
Turn off the unit at the end of the Self test, restore the jumper X21 and switch
on the unit again;
TROLL type pc boards
Press the red button at the end of the Self test, turn off the unit, restore
the jumper X21 and switch on the unit again.
7. Perform a sterilization cycle.
X21
GAM
TROLL
pulsante
rosso
X21
red push-
button
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...