Rev. 1
Page 3-
20
E 041
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
FILLING
PROBLEM
Automatic tank filling
too frequent.
Two automatic fillings every 2
cycles.
Check for water in the external tank.
Fill the tank and switch on the sterilizer to enable the
automatic water filling.
Check the correct pipe connection between
tank and sterilizer.
Restore the connection and check the correct automatic filling
operation.
Check for possible break of the pipe
connecting tank and sterilizer.
Replace the external pipe
Check that the entry filter is not dirty.
Replace the entry filter
Check for possible break of the internal
pipe.
Replace the internal pipe
Check that the water pump works regularly.
Replace the fuse - see card Gr1-16
Check for possible failure of the water
pump.
Replace the water pump- see card Gr3-1
Check the float (minimum level)
Replace the float (minimum level) - see card Gr4-2
Check the integrity of the pipe from the
Milldrop device
Restore the pipe.
Check the START/STOP cable connection
between sterilizer and MILLDROP.
Restore the connection.
Replace the START/STOP cable.
Check the MILLDROP operation.
Perform the maintenance on the Milldrop device –
See Operating Manual
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...