Steam discharge
At the end of the sterilization phase, the steam is released from the sterilization chamber (
STEAM DISCHARGE
). The liquid crystal display
shows:
1
1 3 4 ° C P O R O U S
D E P R E S S U R I Z A T I O N
1 2 3 . 9 ° C
1 . 2 4 b a r 1 8 : 2 0
The icon for the sterilization process status
is
steady on
.
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
from +2,15/1,12 to +0,10 bar
ON
OFF
ON
ON
OFF
ON
ON
OFF
ON
from
+0,10
to
+0,10
bar
ON OFF ON OFF OFF ON ON OFF ON
Steam discharge
from
+0,20
to
-0,00
bar
ON OFF OFF OFF ON ON ON OFF ON