Rev. 1
Page 3-
5
Once the door is open, the user is finally asked to
reset
the system:
( A l a r m M e s s a g e )
R E S E T S Y S T E M
9 5 . 5 ° C X X X X
0 . 0 0 b a r 1 1 : 4 3
Perform a
RESET
(described below) and then turn-off the equipment and check the error or make the repair.
e
When the door is opened, the report (normal or extended depending on the type of alarm) will be printed for the interrupted sterilization
program and the alarm that intervened. Check the document, initial it in the space provided and file it in a suitable place. Refer to
report examples
.
Alarm outside the cycle
If the alarm intervenes
outside the sterilization or test program
the display will show:
( A l a r m M e s s a g e )
A L A R M !
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Turn-off the equipment and check the alarm.
Or, depending on the type of alarm:
( A l a r m M e s s a g e )
W A R N I N G !
8 0 . 5 ° C X X X X
0 . 0 0 b a r
Depending on the alarm
Alarm code
Alarm code
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...