Rev. 1
Page 3-
60
H 401
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
T/P
PROBLEM
Ratio T/P
conv
not
correctly balanced (T
greater than P
conv
)
during the process
phase
The difference value between
P
conv
and T is greater of 2°C
Problem in the board
Replace the CPU board - see card Gr1-1
Problem in the water pump of the steam
generator
Replace the water pump - see card Gr3-1
EV6 valve does not open correctly
Replace EV6 - see card Gr3-1
Replace the lower section of the steam generator
- see card Gr4-6
Steam generator clogged
Different distilled water must be used
Stored calibration values changed
Send the sterilizer to the Mocom Service department
Clean the gasket
Steam leakage from the door gasket
Replace the door gasket- see card Gr6-1
Pressure transducer broken
Replace the pressure transducer - see card Gr1-4
Check the type and the mass of the load
arranged in the chamber
Advice the user on loading and arranging the
material in the chamber as indicated on the
Operating Manual
Check the PT1
Replace PT1 – see card Gr1-8
Calibrate PT1 – see Attachment N
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...