
Rev. 1
Page 3-
40
A 114
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
PT4
SHORTCIRCUIT
Short-circuit in the PT4
probe
The PT4 probe reads a
temperature lower than 1°C
Unstable connection of the PT4 wiring.
Restore the wiring connection.
PT4 wiring out from the board connector.
Restore the wiring connection on the board.
PT4 probe in short-circuit
Replace the PT4 probe - see card Gr1-11
PT4 probe uncalibrated
Perform the calibration – see Attachment N
Too low environmental temperature in the
installation or storage site.
Move the sterilizer in a different installation or storage site
with higher environmental temperature.
Check the wiring of the level sensor for the
external used water tank.
Disconnect the wiring on the CPU board
Check the ground connection of the
sterilizer.
Clean the ground connections.
Replace the damage ground connection.
Sterilizer connected through a power strip.
Remove the power strip and connect the sterilizer as
recommended (see Operating Manual).
Check the electrical system of the room
Conform the ground connection to the standard.
Conform the electrical system to the standard.
Possible data memory fail.
Perform CPU board calibration – see Attachment N
Replace CPU board – see card Gr1-1
Data memory reset.
Replace CPU board – see card Gr1-1 -
CPU board replaced without performing the
calibration.
Perform CPU board calibration – see Attachment N
Replace CPU board – see card Gr1-1
Lost of calibration
Perform CPU board calibration – see Attachment N
Replace CPU board – see card Gr1-1 -
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...