Rev. 1
Page 3-
28
A 024
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
Replace the fuse - see card Gr1-16
LOCKING
PROBLEM
Failure on the micro-
switches (ON-OFF).
During the initial self test the
closing micro-switch is ON
and the opening micro-switch
is OFF
Locking mechanism fuse burned during a
cycle
Go to Attachment H or Attachment I.
Go to Attachment H or Attachment I
Alarm occurred during the previous cycle;
sterilizer turned off before completing the
safety procedure
Advice the user for the correct operation in case of alarm
Replace the motor - see card Gr6-6
Locking mechanism broken
Replace the motor coupling pin - see card Gr6-6
Burned fuse.
Replace the fuse – see card Gr1-16.