LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
PT1 BROKEN
PT1 faulty
The temperature detected by
the PT1 is higher than 250°C
PT1 broken
Replace the PT1 - see card Gr1-8
PT1 uncalibrated.
Perform the calibration – see Attachment N
Check the temperature value on the display.
If = the room temperature:
- turn off and on the sterilizer more times;
- Replace PT1 – see card Gr1-8.
If differs for many °C:
- turn off and on the sterilizer more times;
- Replace PT1 – see card Gr1-8;
- Reset and calibrate the CPU board – see
Attachment N.
Check the calibration.
Perform the PT1 calibration– see Attachment M.
Complete sterilizer calibration – see Attachment N
Replace the CPU board – see card Gr1-1
Reset or failure on the data memory.
Reset and calibrate CPU board – see Attachment N
Replace the CPU board – see card Gr1-1
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...