Filling distilled water
Manual filling
The first time the sterilizer is used, and later when the MIN water level indicator comes on, you will have to fill, or top-off, the internal distilled
water tank.
With reference to the figure (and with the door open), proceed as follows:
1. Fill the manual container (2 l) with distilled water, keeping it horizontal;
2. Connect the tube’s quick connector to the corresponding female connector under the chamber entrance (marked
), pushing
until you hear a click;
3. Place the container in a vertical position, at the same time, loosening the plug and taking care not to spill water on the machine.
4. The water will begin to flow into the tank;
5. Continue filling until the MIN level indicator
turns off.
6. Continue until the water is drained from the container;
7. At this point, lower the connector below the connection point, keeping it horizontal;
8. While pinching the tube with your fingers, press the metal lever located on the side of the connector and detach the quick connector;
9. Refill the container (2 l) and repeat the operations described in points 2, 3 and 4 a second time;
10. When the MAX level icon comes on (accompanied by a beep), stop filling and detach the quick connector as described in points 7 and
8.
e
The icon MAX does not have to be on to start a sterilization program. The icon
MIN
indicator off is
sufficient
.
e
Use ONLY high quality distilled water. For the specifications of the water supply, see
Technical Characteristics
.
Automatic filling
In the event of sterilzer installation for automatic filling from an external tank or demineralizer Milldrop (see “
Installation
”
), the filling will occur
automatically after the automatic filling option has been selected. Obviously, for the correct operation, the user must fill the external tank
or switch
on the Milldrop in advance.
To set the automatic filling option, please refer to the
Chapter “
Filling Option
”
.
T
HE AUTOMATICALLY FILLING SYSTEM MUST NEVER RUN DRY
;
THIS CAUSES PREMATURE WEAR TO THE AUXILIARY WATER
-
INJECTION PUMP
.
P
ERIODICALLY CHECK THE
WATER LEVEL IN THE EXTERNAL TANK
.
REMOVING THE TUBE
TANK FULL
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...