Rev. 1
Page 3-
64
H 405
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
P OVER LIMIT
Pressure value over
the MAX threshold
during the process
phase
Pressure value greater than
1.24 or 2.31 bar (depending
on the cycle type: 121 or
134°C)
Fault in the board
Replace the CPU board - see card Gr1-1
Stored calibration values changed
Send the sterilizer to the Mocom Service department
Clean the valve
Discharge valve does not open correctly
Replace the valve - see cards Gr2
Pressure transducer uncalibrated
Send the sterilizer to the Mocom Service department
Pressure transducer broken
Replace the pressure transducer - see card Gr1-4