Rev. 1
Page 3-
30
A 040
LCD INDICATION
Alarm description
Effect/Signaling
Possible cause / Check
Proposed solution
FILLING
PROBLEM
Fail of the automatic
filling from the
external tank.
The MlN level signaling is not
turned off within 2’ from the
start command of the
automatic filling.
Check for water into the external filling tank.
Fill the external tank, reset the alarm and let the sterilizer’s
main tank to be filled.
Check that the tap in the external filling tank
is in open position.
Open the tap in the external filling tank.
Check for possible obstruction in the
external filling tank pipe.
Free the external filling tank pipe and shorten it if too long.
Check for possible dirty in the Saeco filter.
Replace the filter in the filling pipe path
Replace the fuse - see card Gr1-16
Replace the water pump wiring
Replace the water pump - see card Gr3-1
Verify that the water pump works properly.
Replace the power board - see card Gr1-1
Beat on the pump body to help the start of water pump.
Verify that the water pump is not blocked
Replace the water pump - see card Gr3-1
Fault of the water pump
Replace the water pump - see card Gr3-1
Continue
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...