ATTACHMENT J -
RECOVER DEFAULT DATA
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove jumper X21 of the CPU board;
3. Push on the three left keys and switch-on the unit;
4. LCD appears completely empty;
5. Switch-off and -on again the unit;
6. LCD will show now the message “DEFAULT DATA UPDATE”;
7. Wait for completing the update with the message “OK”;
8. Switch-off the unit;
9. Restore the jumper X21 and switch on again the unit;
10. Perform a sterilization cycle.
WARNING
As consequence of the data recovery operation, the sterilizer must be calibrated;
this procedure should be performed starting from the software release
E0008 /
BP00350