Rev. 1
Card Gr1-
19
BAND CHAMBER HEATER
410#00140000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card
Gr7-1
);
2. Remove the reservoir assembly – see card
G4-1
;
3. Unscrew the central rail (1);
4. Remove the thermostat probe (2) from the heater – see card
Gr1-13
;
5. Remove PT3 probe (3) – see card
Gr1-10
;
6. Remove the insulation layer covering the chamber heater;
7. Disconnect the heater wiring from the power board;
8. Remove the screws (4) fixing the heater (5) and withdraw it carefully out the
chamber surface;
9. Mount the new chamber heater, restore connections and assembly all items
proceeding in reverse order as above;
10. Run a sterilization cycle.