SAFETY DEVICES
The sterilizer is equipped with the following safety devices for which we provide a
brief description of their function:
–
Mains fuses
(see summary table data)
Protection inside the device against a fault in the heating elements.
Action: cuts the electricity.
–
Fuses protecting the electronic circuits
(
see summary table data
)
Protection against a fault in the primary transformer circuit and low voltage uses.
Action: cuts power to one or more low-voltage circuits.
–
Thermal circuit breakers on the mains voltage windings
Protection against overheating of the vacuum pump motor and the primary transformer
windings.
Action: temporary cut-off (until cooling) of the winding.
–
Safety valve
Protection against overpressure in the sterilization chamber.
Action: release of the steam and restoration of the safety pressure.
–
Steam generator manual rearm safety thermostat
Protection against steam generator overheating.
Action: cut-off of the electricity to the steam generator.
–
Heating element manual rearm safety thermostat
Protection against overheating of the heating elements of the container under pressure.
Action: cut-off of the electricity to the chamber heating element.
–
Door position safety microswitch
Confirmation of the correct closing position of the door of the container under pressure.
Action: signals wrong door position.
–
Mechanized door lock mechanism with electromechanical protection (pressure
switch)
Protection against accidental opening of the door (even in a blackout).
Action: prevents accidental opening of the door during a program.
–
Door lock mechanism safety microswitch
Confirmation of the correct closing of the door lock.
Action: signaling the failure or incorrect operation of the door lock mechanism.
–
Self-leveling plumbing system
Plumbing system structure for the spontaneous leveling of the pressure in the case of a
manual interruption of the cycle, alarm or blackout.
Action: automatic restoration of atmospheric pressure in the sterilization chamber.
–
Integrated system for evaluating the sterilization process
Continuous verification of the sterilization process parameters entirely managed by
microprocessor.
Action: immediate interruption of the program (in case of anomaly) and generation of
alarms.
–
Monitoring of the sterilizer's operation
Real-time oversight of all significant parameters when the machine is powered.
Action: generation of alarm messages (in the case of anomaly) with possible interruption
of the cycle.
Содержание millennium B
Страница 1: ...Rev 1...
Страница 202: ...Rev 1 Gr2 INDEX GROUP 2 ELECTROVALVES EV1 VALVE 1 EV2 VALVE 2 EV3 VALVE 3 EV4 VALVE 4 EV5 VALVE 5...
Страница 208: ...Rev 1 Gr3 INDEX GROUP 3 PUMPS STEAM GENERATOR WATER PUMP 1 AUTOMATIC WATER FILLING PUMP 2 VACUUM DOUBLE PUMP 3...
Страница 243: ...Rev 1 Gr7 INDEX GROUP 7 COVERS FRAME COVER 1 DOOR COVER 2 FRONT FRAME 3 SERVICE DOOR 4...
Страница 270: ...Rev 1 Page 4 4 HYDRAULIC DIAGRAM CHAMBER HEAT EXCHANGER WP VP Main tank...
Страница 271: ...Rev 1 Page 4 5 EXPLODED VIEWS TABLE 1...
Страница 272: ...Rev 1 Page 4 6 TABLE 2...
Страница 273: ...Rev 1 Page 4 7 TABLE 3...
Страница 274: ...Rev 1 Page 4 8 TABLE 4...
Страница 275: ...Rev 1 Page 4 9 TABLE 5...
Страница 276: ...Rev 1 Page 4 10 TABLE 6...