77
Cinemage User Manual
Version 4.1.x ©2010
Steps to calibrate your display: Full Gamut
[Option]
Before navigating to the Display Setup menu plug the GretagMacbeth (GMB) calibration probe into the
USB port. If the probe is not plugged in before navigating to this menu, the monitor will not be able to
“see” the probe until the unit is restarted with the probe attached.
1. From the main menu press select “Setup Menus”
2. Select Video &Display Setup.
3. Select More Display Setup.
4. Select Measure (note: do not select Auto Calibrate yet).
5. If you are using the GMB EyeOne Pro place the probe on the white calibration tile provided
with the probe and Select “Read CIE xyY” and follow the on screen directions. (on screen
directions).Then place probe on the center of the display. If you are using the EyeOne
Display 2 place the probe on a flat black surface (the top of the monitor works well) then
press “Read CIE xyY”. When finished place the probe on the center of the display.
6. Press “Flat Field” until you have “100% White” displayed.
7. Adjust the backlight to a comfortable setting for your current ambient environment.
8. Press “Measure Now”.
Read the results for display luminance out “Y” under “CIE x,y,Y” Adjust the backlight either up
or down and repeat step 7-8 until you reached your desired luminance output setting. The
readout will be in both Candelas and Foot Lamberts.
9. Press the back button twice to return to the Video & Display Setup Menu
10. Select “Monitor Profiling”
11. Hit the “Change Sample Count” to determine how accurate of a profile you want to perform
on your display. The larger the sample count, the more accurate the profile will be. The
tradeoff however is that the larger the sample count, the longer the profile will take to
complete. After selecting the Sample Count size, hit the back button. The recommended
size is 64.
12. Press the “Auto Profile Display” button and follow the on screen directions as they are given
to you.
13. (On screen directions) If you are using a GretagMacbeth
EyeOne Display2
place the probe
on a flat black surface (the top of the monitor works well) then press “press to calibrate
probe”. When finished place the probe on the center of the display. If you are using the
GretagMacbeth
EyeOneDesign
place the probe on the calibration tile provided with the Eye
One Design and Press “Place Probe and Press”. Use the LCD mounting device provided with
the Eye One Design and mount the probe on the middle of the display.
14. Check the results and make adjustments to the backlight for Luminance level, and Manual
Calibration settings as needed until you have the desired results.
Note:
There is no need to
perform another Auto Profile of the monitor after making manual calibration adjustment.
Performing another Auto Profile may cause unintended effects due to the nature of the
calibration process.
15. Save the results of your manual calibration changes.
16. Calibration Complete.