44
Cinemage User Manual
Version 4.1.x ©2010
Calibration:
Before navigating the menus, plug the GretagMacbeth (GMB) calibration probe
[option]
into the USB port. From the main menu, press the Setup Menus button:
↑
Then select the Video & Display Setup button:
↑
Then select the More Display Setup button:
↑
Then select the Measure button (note: do not select Auto Calibrate yet):
↑
If you are using the GMB EyeOne Pro place the probe on the white calibration tile
provided with the probe and Select “Press to Calibrate Probe” Then place the probe on
the center of the display. If you are using the EyeOne Display 2 place the probe on a flat
black surface (the top of the monitor works well) then press “press to calibrate probe”.
When finished place the probe on the center of the display. Press the Set Flat Field
button until “100% White” is displayed:
↑
Cinemage 2142
Press to lockout menus
Operator Menus
System Menus
Setup Menus
Display Controls
Presets
Back
Unit Information
Video & Display
Setup
Preset & Preferences
Setup
Network & USB
Datakey Setup
Reset
Back
Video Setup
Calibration
Setting
Calibration
Type:
Monitor
Profiling
More Display
Setup
Backlight
Brightness
Back
Adjust Calibration
Settings
Profiling Options
Manual Calibration
Adjustments
Blue Only
Mode
Meausre
Backlight
Brightness
Back
Hold and release for
probe info
Read CIE xyY
x=.314, y=.345, Y=115Cd
Read CIE XYZ
X=.314, Y=.345, Y=115 Cd
Select Calibration
Video Source
Set Flat Field
Backlight Brightness
20.0%