Troubleshooting
7-7
Clean water Raman scan.
0
1
2
3
4
5
365
380
395
410
425
440
455
Wavelength (nm)
Intensity (10
5
counts/s)
397 nm
C o n ta m in a te d w a te r in a w a te r R a m a n s c a n .
0
1
2
3
4
5
3 6 5
3 8 0
3 9 5
4 1 0
4 2 5
4 4 0
4 5 5
W a v e le ngth (nm )
Intensity (10
5
counts/s)
Contaminated water or dirty
cuvette (note high background
)
Clean water
Water Raman spectra
Contaminated water
Running a water Raman scan helps identify abnormalities caused by accessory prob-
lems or miscalibration. The following spectrum is normal:
Below is a normal water Raman spectrum superimposed on one that exhibits a problem.
In this instance, the water was contaminated, resulting in a high background.